K600
6.2.1 PARTIAL RESET
To reset the Partial Register, finish dispensing and wait for the
meter to show a Flow Rate of 0.0 as indicated in the illustration
then quickly press RESET
NOTE
Unlike Normal mode, in this case during reset, you do not pass through the stages
where the display segments are first lit up and then switched off, but rather the
reset partial register is immediately displayed
7 CALIBRATION
7.1 DEFINITIONS
Calibration
factor or “K
Factor”
This is the multiplication factor applied by the system to the electrical pulses received, to transform
these into measured fluid units
Factory K
Factor
Factory-set default factor. It is equal to 1,000.
This calibration factor ensures utmost precision in the following operating conditions:
Version for
diesel fuel
Fluid Diesel
Temperature : 38°C
Flow rate : 10-100 litres/min
Version for oil Fluid motor oil type SAE10W40
Temperature : 20°C
Flow rate : 6-60 litres/minn
USER K FAC-
TOR
Customized calibration factor, meaning modified by calibration.
Even after any changes have been made by the user, the factory K factor can be restored by means of a simple procedure.
7.2 CALIBRATION MODE
Why calibrate
K600 METER is supplied with a factory calibration that ensures precise measuring in
most operating conditions.
Nevertheless, when operating close to extreme conditions, such as for instance:
-
with fluids close to acceptable range extremes (such as low-viscosity antifreeze or high-
viscosity oils for gearboxes)
-
in extreme flow rate conditions (close to minimum or maximum acceptable values) on-the-
spot calibration may be required to suit the real conditions in which the meter is required
to operate.
When operating close to extreme use or flow rate conditions (close to minimum or maxi-
mum acceptable values), an on-the-spot calibration may be required to suit the real condi-
tions in which the K600 is required to operate.
FOREWORD
K600 METER permits making quick and precise electronic calibration by changing the
Calibration Factor (K FACTOR).
Two procedures are available for changing the Calibration Factor:
1 In-Field Calibration, performed by means of a dispensing operation
2
Direct Calibration, performed by directly changing the calibration factor
The calibration phases can be entered (by keeping the CAL key pressed for a long time) to :
-
Display the currently used calibration factor