9.2 CLEANING
CLEANING THE
MEASURING
CHAMBER
The K600 measuring chamber can be cleaned without removing the instru-
ment from the line on which it is fitted. Make sure the gears are turning freely
before closing the cover.
ATTENZIONE Always make sure that the liquid has drained from the meter before
cleaning.
To clean the chamber, proceed as follows (with reference to the exploded diagram
positions):
1
Loosen the four retention screws of the lower cover (pos. 7).
2
Remove the cover (pos. 7) and the seal (pos. 6).
3
Remove the oval gears.
4
Clean where necessary. For this operation, use a brush or pointed object such
as a small screwdriver.
Be careful not to damage the body or the gears.
5
To reassemble the instrument, perform the operations in the opposite se-
quence.
ATTENTION Perform the assembly diagram to reassemble the gears.
Only one of the two gears, modularly coupled as shown in the picture
aside, features magnets. Observe the position of the gear with mag-
nets, as shown in the figure. Fit the second gear (without magnets) with
axis greater than 90° compared to the first gear.
CLEANING THE
FILTER
The filter cleaning interval is to be defined depending on the impurities con-
tained in the fluid. To perform this operation, remove the device from the line
on which it is installed, as the filter is placed between the meter body and tube
connection flange.
ATTENTION Always make sure that the liquid has drained from the meter before
cleaning.
To clean the filter, proceed as follows (with reference to the exploded diagram positions):
1
To access the filtering disk of the K600/3, loosen the 2 fixing screws of the con-
nection flange at the inlet. Remove both flanges if it is necessary for the system.
2
Remove the meter from the line, being careful to remove also the gaskets be-
tween the flanges and threaded connections of K600.
3
Slide out the filter (pos. 9)
4
Clean the filter with compressed air.
5
Carry out the reverse procedure to reassemble the filter.