EasyManua.ls Logo

+GF+ Signet 4630 - Relay Settings

+GF+ Signet 4630
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
27
463X Chlorine Systems Manual
Example: Set a relay to energize at a low setpoint of
1.0 ppm with a time delay of 15 seconds and de-energize
at 1.30 ppm.
Once a setting is saved it immediately becomes
active.
1. Go to the Calibrate Menu.
2. Scroll down to the Relay 1 Mode menu.
3. Press to enter Edit Mode.
4. Scroll down and choose Low.
5. Press Enter.
6. Scroll down to the Relay 1 Source menu. The default
is Free Chlorine.
7. Scroll down to the Relay 1 Setpnt menu.
8. Press to enter Edit Mode.
9. Set the ppm value to trigger the relay: 1.00 ppm.
10. Press Enter.
11. Scroll down to the Relay 1 Hys menu.
12. Press to enter Edit Mode.
13. Set the hysteresis (dead zone) for this relay. This affects
the turn off only: 0.3 ppm.
14. Press Enter.
15. Scroll down to the Relay 1 Delay menu.
16. Press to enter Edit Mode.
17. Set the turn-on delay in seconds for the relay: 15 secs.
18. Press Enter.
19. Exit to View Mode.
Relay function can be tested in the Options Menu.
EXAMPLE
Relay Settings
Relay energized
Relay de-energized
1.00
1.30
Time
Cl ppm
Time Delay
Low
Setpoint
Calibrate Menu
Step 4
6
9
13
17
Set and save
Relay 1 Mode:
Low >
Set and save
Relay 1 Source:
Chlorine >
Set and save
Relay 1 Setpnt:
1.00 ppm >
Set and save
Relay 1 Hys:
0.30 ppm >
Set and save
Relay 1 Delay:
15.0 secs >
EXAMPLE
Output Settings - Loops and Relays

Related product manuals