28
463X Chlorine Systems Manual
CAUTION!
Remove power before wiring.
Follow instructions carefully to avoid personal injury or
damage to the electronics.
Refer to the wiring diagram above to connect the 2650 cable to the terminals on the 8630 Chlorine Transmitter.
For calibration and con guration please refer to the 8630 Calibrate Menu discussions above.
The electronics are pre-wired from the factory to the transmitter. Refer to the following
schematics when replacing the electronics.
8630
Chlorine
Gnd
I/O
V
+
18
17
16
Signal Ground
Digital (S
3
L) data
+5 VDC
White
Red
Black
Shield
pH
Signet 3-2750-7
(159 001 671)
N/C
8630
Chlorine
Gnd
I/O
V
+
15
14
13
Signal Ground
Digital (S
3
L) data
+5 VDC
White
Red
Black
Shield
Chlorine
Signet 3-2650-7
(159 001 670)
N/C
The Signet 2650-7 Amperometric Electronics provide the polarization voltage and signal conditioning required by the Signet
2630-X and 2632-X Sensors.
The Signet 2750-7 pH Electronics conditions and ampli es the output from the Signet 2724-00 pH Electrode.
Both units output a Digital (S
3
L) signal to the Signet 8630 Chlorine Transmitter.
2650-7 Amperometric and 2750-7 pH DryLoc
®
Electronics
Wiring to the Signet 8630 Chlorine Transmitter