7
463X Chlorine Systems Manual
NO
Bright light can promote algae growth. Indoor mounting is recommended.
If the system is mounted outdoors, an outdoor enclosure for the whole system must be
used to protect the electronics and ow cell from light, rodents, insects and dirt.
Mount the panel according to local electrical, building, and/or plumbing codes and
seismic requirements.
Use four 6 mm (¼ in.) diameter screws or bolts of suf cient length to mount the panel to
a sturdy vertical surface. A mounting template is provided.
Allow clearance on the sides and bottom for service to the unit.
Keep panel system electronics and enclosure away from dripping water.
The panel must be mounted vertically in an upright position.
Do not mount in direct sunlight.