EasyManua.ls Logo

+GF+ Signet 4630 - Troubleshooting - 463 X Chlorine Analyzer

+GF+ Signet 4630
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
39
463X Chlorine Systems Manual
The troubleshooting table below outlines possible causes and remedies related to the ow cell panel system.
Refer to the respective transmitter and sensor manuals for speci c component troubleshooting.
Problem Possible Cause Remedies
No water ow
Inlet pressure below 1 bar (15 psi)
Check source pressure. Remove regulator if pressure is
< 1 bar/15 psi
Damaged ow regulator Replace ow regulator
Clogged lter
Clean lter or remove pressure regulator if inlet pressure is
below 1 bar (15 psi)
Low water ow
Clogged lter Clean lter
Insuf cient inlet pressure
Increase inlet pressure to speci cation or remove ow
regulator
Damaged ow regulator Inspect or replace
Excessive ow
Excessive pressure over 8 bar (120 psi) Ensure inlet pressure does not exceed 8 bar (120 psi)
Damaged ow regulator Replace ow regulator
Flow cell leaks
Incorrect assembly Inspect, clean, and reassemble
Loose bolts Tighten bolts (8.1 Nm, 72 Lb-In Max)
Defective or missing O-rings Replace O-rings
Damaged sealing surfaces Replace ow cell
Leaks around sensor
Damaged sensor O-ring Inspect or replace O-ring
Damaged ow cell Inspect ow cell
Water leaks out of top vent hole
Drain valve closed or plugged Inspect
Hose connections are backwards Connect water source correctly
Outlet drain not below ow cell Route drain line below ow cell and vented to atmosphere
Algae growth in ow cell
Low chlorine concentration in water
Locate ow cell away from light
Exposure to light
Excessive small bubbles in ow
cell accumulating on sensors
Water source contains bubbles
Check water source
Water source is saturated with
dissolved gasses
Transmitter does not turn on
No power or external circuit breaker is
switched off
Inspect power and wiring connections
Loose or incorrect wire connections Check connections
Open loop wiring or missing jumper
wire on terminal 1 (Loop 1) on the
wiring enclosure terminal block
Replace jumper wire to terminal 1 or connect a loop device
if the jumper wire to terminal 1 is removed
Chlorine reading too high Breached membrane Replace membrane and ll solution
Chlorine reading too low
Electrode needs to be repolished Polish gold electrode
Not enough electrolyte Check electrolyte level
Low ow rate Check source pressure
Filter clogged Clean the lter
Sensor conditioning not long enough
Allow the sensor to condition for 4 hours
(See 2630 or 2632 sensor manual for details)
Sensor output drifts
Electrode needs to be repolished Polish gold electrode
Not enough electrolyte Check electrolyte level
Variable ow rate Check ow rate
Clogged lter Clean the lter
Contaminated ll solution Change electrolyte ll solution
Chlorine Electrode Troubleshooting (see 2630-2 or 2632-1 electrode manual: 3-2630.090)
Transmitter Troubleshooting (see 8630-3 transmitter manual: 3-8630.090-3)
pH Electrode Troubleshooting (see 2724 electrode manual: 3-2724.090)
Troubleshooting - 463X Chlorine Analyzer
Appendix

Related product manuals