47
463X Chlorine Systems Manual
General
Compatibility: Signet 2750 pH/ORP Sensor electronics
Process
Connection: ¾ in. NPT or ISO 7/1 R¾ threads
or Signet ow ttings
Wetted Materials: PPS, glass, UHMW PE, FPM
Performance
Ef ciency: >97% @ 25 °C (77 °F)
Ef ciency indicates the "wellness" of a
new electrode. Ef ciency is measured by
comparing the actual slope (mV/pH) at 25
°C to the theoretical output of 59.16 mV/pH.
An ef ciency of 97% to 100% is equivalent
to a slope of 57.39 to 59.16 mV/pH.
Range: 0 to 14 pH
Environmental
Pressure
Operating:
-10 ºC to 65 °C (14 °F to 149 °F),0 to 6.9 bar (0 to 100 psi)
65 °C to 85 °C (149 °F to 185 °F),
Linearity derated 6.9 to 4.0 bar (100 to 58 psi)
Temperature
Storage: 0 °C to 50 °C (32 °F to 120 °F).
CAUTION: The electrode glass will
be shattered if shipped or stored at
temperatures below 0 °C (32 °F).
The performance life of the electrode will
be shortened if stored at temperatures
above 50 °C (120 °F).
Operating: -10 ºC to 85 ºC (14 ºF to 185 ºF)
Rel. Humidity: 0 to 95% indoor/outdoor non-condensing to
rated ambient
Shipping Weight: 0.25 kg (0.55 lb)
Standards and Approvals
WEEE
RoHS Compliant
Manufactured under ISO 9001 for Quality,
ISO 14001 for Environmental Management and
OHSAS 18001 for occupational health and safety.
China RoHS (Go to www.gfsignet.com for details)