Eureka3 User Manual
Eureka3 Page 11
Whilst charging the LEDs will show the status as below:
Normal charging Green Charge level display (1-5 LEDs)
Over temperature Flashing Green Charge level display (1-5 LEDs)
Charging complete Green All on
Fault Red Off
The over temperature state may occur when charging in high temperatures. The transmitter
electronics will manage this state and continue to charge when possible.
The fault state indicates a problem and the transmitter should be returned to a Primayer Service
Centre for further investigation.
Only Eureka3 charge leads should be used to
charge Eureka3 transmitters. DO NOT attempt
to use Eureka2 or Eureka Digital charge leads,
as damage may occur which will not be covered
by warranty.
3.1.7.3.1 Charge Time
Charge time from flat to full is between 3 hours 45 minutes and 4 hours 30 minutes. The charge time
will be longer in higher temperatures; there is automatic monitoring in the electronics to prevent the
batteries overheating.
3.1.7.3.2 Battery Life
Transmitter battery life is up to 18 hours, depending on usage pattern.
3.1.8 Internal Accelerometer
The internal accelerometer is mounted on the base of the transmitter and is magnetic. A stability plate
can be magnetically mounted to increase the area of the base and provide greater stability for use
when the internal accelerometer is not used.