Important Control Solution Information
Use only Prodigy
®
Control Solutions.
Check the expiration date on the control solution bottle. Do not use if expired.
Control solution, meter, and test strips should come to room temperature
(68-77º F/20-2 C) before testing.
Shake the bottle before use, discard the first drop of control solution after squeezing,
wipe off the dispenser tip to avoid contanimation. These steps ensure you will get a good
sample and an accurate result.
Use within a period of 90 dyas from the date opened. Record the discard date (date opened
plus 90 days) on the control solution bottle. Discard after 90 dyas.
Store the control solution tightly closed at temperatures below 86ºF (30ºC).
Do not refrigerate.
PLEASE NOTE
The control solution range printed on the test strip vial is for Prodigy
®
control solutions only.
It is used to test meter and test strip performance. It is not recommended range for your
blood glucose level.