Blood Glucose Monitoring System - User’s Manual
Diagnostic Devices, Inc www.prodigymeter.com 1.800.243.2636
19
• Important Control Solution Information
• Use only Prodigy
®
Control Solutions.
• Check the expiration date on the control solution bottle. Do not use if expired.
• Control solution, meter, and test strips should come to room temperature
(68-77º F/20-25º C) before testing.
• Shake the bottle before use, discard the first drop of control solution after squeezing,
wipe off the dispenser tip to avoid contanimation. These steps ensure you will get a good
sample and an accurate result.
• Use within a period of 90 dyas from the date opened. Record the discard date (date opened
plus 90 days) on the control solution bottle. Discard after 90 dyas.
• Store the control solution tightly closed at temperatures below 86ºF (30ºC).
Do not refrigerate.
PLEASE NOTE
The control solution range printed on the test strip vial is for Prodigy
®
control solutions only.
It is used to test meter and test strip performance. It is not recommended range for your
blood glucose level.