Blood Glucose Monitoring System - User’s Manual
Diagnostic Devices, Inc www.prodigymeter.com 1.800.243.2636
43
Error Message
Battery is dead.
Please replace.
MESSAGE
WHAT DOES WHAT IT
ACTION
METER SAY? MEANS
Strip has been used.
Please replace.
The room
temperature is out
of range.
Thus unable to
measure.
Appears when the
batteries cannot
provide enough
power for a test.
Replace the batteries
immediately.
Appears when
inserting a used test
strip.
Test with a new
test strip.
Appears when
ambient temperature
is below system
operation range.
50º F (10º C)
Appears when
ambient temperature
is above system
operation range.
104º F (40º C)
System operation
range is 50ºF~104ºF
(10ºC~40ºC). Repeat
the test after the
meter and test strip
have reached the
above temperature.