Error message: Other fault
Confirm error message with ENTER key
Possible cause External fault signalling device triggered
Remedy Rectify external cause of fault
Possible cause No external fault signalling device connected and no jumper is connected across the contacts at
the fault input
Remedy Connect jumper across contact at fault input
Error message: Mains voltage too low
Cause The supplied voltage is or was lower than the minimum permissible supply voltage
Remedy Check supply voltage
Error message: Bus fault
Possible cause Bus connection interrupted
Remedy Re-establish bus connection (see Function and Fault Indicators on Ballast Units)
Possible cause Power supply not sufficient on one or several ballasts
Remedy Check power supply or fine fuse in ballast
(only by authorised electrician!)
Possible cause Ballast defective
Remedy Replace ballast (only by authorised electrician!)
Error message: Memory error
Cause The control determined an error in the memory as part of the self-test
Remedy Replace control unit (only by authorised electrician!)
Error message: Basic setting
Cause The control determined an error as part of the self-test
Remedy Replace control unit (only by authorised electrician!)
Error message: Lamp current
Cause Lamp current set outside the permissible limits
Remedy Set lamp current within permissible limits (see datasheet)