Page 42
8.3 Electrical Data
Radiation chamber
Lamp type: OptiFlux 300 W
Lamp current: Standard: 3.5 A
Maximum: 3.5 A
Minimum: 3.0 A
Control
Nominal voltage: 230 V AC, 50/60 Hz
Fuses: Fine fuse 5 x 20 mm (250 V AC):
Upper fuse (control power supply):
0.16 A slow-blow, Order No. 712048
Lower fuse (switched power outputs (XR1-XR3)):
2.5 A slow-blow, Order No. 712033
NOTE
The fuses are located in fuse holders with bayonet catches on the right in the terminal box
of the control system.
Control cabinet
Inputs
• Contact inputs (-X3:1 ... -X3.6):
for contacts or switching transistors:
No-load voltage: 5V ± 0.5 V
Input resistance: 10 kOhm
Outputs
• Voltage outputs (-X1:1 ... -X1.10):
Type of contact: Normally open contact
Load capacity: 250 V AC / 3 A / 100 VA
Provide RC protective circuits for inductive loads!
• Relay outputs (-X4:1 ... -X4.6):
Type of contact: Normally open contact
Load capacity: 250 V AC / 3 A / 100 VA
Provide RC protective circuits for inductive loads!
• Alarm relays (-X4:7 ... -X4.9):
Type of contact: Changeover contact
Load capacity: 250 V AC / 3 A / 100 VA
Provide RC protective circuits for inductive loads!
• Standard signal output mA (-X3:7 ... -X3.8):
0/4...20 mA, floating
Maximum load: 600 Ohm
Technical Data