Page 46
U-Throated Flume
ABSOLUTE
If the flow calculation is to be absolute P702 = 1 the flow will be calculated
using the formula: q = (2/3)
1.5
g
n
0.5
C
u
C
v
C
d
bh
1.5
Where: q = flowrate
g
n
= gravitational acceleration (nominal value = 980.66 cm/s
2
) h
= head C
u
=
shape coefficient calculated by FlowCERT LITE P724
C
v
= velocity coefficient calculated by FlowCERT LITE P721
C
d
= discharge coeffecient calculated by FlowCERT LITE P722
b = throat width P711
RATIOMETRIC
U-Throated Flume
If the flow calculation is to be ratiometric P702 = 2 the flow will be
calculated using the formula:
q= q
cal
(C
v
/C
vcal
)(C
d
/C
dcal
)(C
u
/C
ucal
)(h/h
cal
)
1.5
Where: q = flowrate
q
cal
= flowrate at maximum head P705
Cv = velocity coefficient calculated by FlowCERT LITE P721
Cv
cal
= velocity coefficient at maximum head
Cd = discharge coeffecient calculated by FlowCERT LITE P722
Cd
cal
= discharge coefficient at maximum head
Cu = shape coefficient P724
Cu
cal
= shape coefficient at maximum head
h = head
h
cal
= maximum head P704

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pulsar FlowCERT LITE and is the answer not in the manual?

Pulsar FlowCERT LITE Specifications

General IconGeneral
BrandPulsar
ModelFlowCERT LITE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals