9 000690871.NA Rev. A Q50 R Carbon
Table 3 – Proximity •elds from RF wireless communica• ons equipment
Test frequency (MHz) Band (MHz)
Immunity test levels
Home healthcare environment
385 380 - 390 Pulse modula• on 18Hz, 27V/m
450 430 - 470 FM, ±5kHz devia• on, 1kHz sine, 28V/m
710
704 - 787 Pulse modula• on 217Hz, 9V/m745
780
810
800 - 960 Pulse modula• on 18Hz, 28V/m870
930
1720
1700 - 1990 Pulse modula• on 217Hz, 28V/m1845
1970
2450 2400 - 2570 Pulse modula• on 217Hz, 28V/m
5240
5100 - 5800 Pulse modula• on 217Hz, 9V/m5500
5785
Table 4 - Input a.c. power Port
Phenomenon Basic EMC standard
Immunity test levels
Home healthcare environment
Electrical fast transients/burst IEC 61000-4-4
±2 kV
100kHz repe• • on frequency
Surges Line-to-line IEC 61000-4-5 ±0.5 kV, ±1 kV
Conducted disturbances induced by RF
•elds
IEC 61000-4-6
0% UT; 0.5 cycle
At 0º, 45º, 90º, 135º, 180º, 225º, 270º
and 315º
0% UT; 1 cycle and
70% UT; 25/30 cycles Single phase: at 0º
Voltage interrup• ons IEC 61000-4-11
0% UT; 250/300 cycles
UT=rated input voltage
Table 5 - Signal input/output parts Port
Phenomenon Compliance Electromagne! c environment
Electrosta• c Discharge IEC 61000-4-2
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air