10.3. Return to Weighing
Introduced modifications are automatically recorded upon return to the home
screen. To return to the home screen press key repeatedly.
11. WEIGHING
Set high filter rate for those patients who can't stay motionless (read section
12.1). With this the weighing will take longer, but the result will be more
reliable.
11.1. Zeroing
To zero mass indication press key. Zero indication and the following
pictograms are displayed: and . The scale can be zeroed only when
the indication is stable.
Indication can be zeroed only within ±2% range of the
maximum capacity. If the zeroed value exceeds ±2% of the
maximum capacity, then the software indicates a respective
error message, Err2.
11.2. Taring
Spread a nappy or a blanket over the weighing pan (this is conditioned by
a scale type). Upon indication stabilization press key (zero indication
is displayed, Net pictogram appears at the top of the screen). Now, depending
on the scale type:
load the weighing pan with a baby (baby scale),
stand on the weighing pan (personal scale),
sit on the chair (chair scale),
lay on the bed (bed scale).
Read the result when stability marker is displayed.
Baby scale is intended for measurement of baby mass, the
baby is weighed lying. Shall the baby be weighed sitting,
make sure that it sits in the centre of the weighing pan.
Otherwise the pan may rest on bumpers protecting against
overload, which might make the measurement unreliable.