Manual printout of a stable weighing result. Upon pressing key at the moment
when the result is unstable (no pictogram displayed), the program first waits for
the stability condition to be met, only then the printout is carried out.
Manual printout of each weighing result. In case of unstable indication, <?> sign is
displayed in front of the 'mass frame'. Function available for non-verified scales
exclusively.
Automatic printout of the first stable weighing result above <Lo> threshold (to set
<Lo> threshold go to <Lo> parameter).
Automatic printout with time interval set in [min] (to set the interval go to <Int>
parameter).
Press key to confirm, next go to the home screen.
18.2.3. Automatic Printout Interval
Parameter enabling to set frequency of an automatic printout. Printout interval
is set in minutes with 1 [min] readability within 1 [min] - 1440 [min] range.
Procedure:
Enter <P3.Func> menu and select a given working mode.
Enter <Int> function, window for entering time interval value is displayed.
Press key to confirm, next go to the home screen.
18.2.4. LO Threshold
<Lo> parameter allows to configure the function of automatic operation.
In order to save the next measurement, before carrying it out, the mass
indication must get below the set net value of Lo threshold.
Procedure:
Enter <P3.Func> menu and select a given working mode.
Enter <Lo> function, window for entering Lo threshold value is displayed.
Enter respective value and press key to confirm, then continue
weighing.
19. WORKING MODE – WEIGHING
<UUGG> is a standard working mode enabling to carry out the weighing
operation along with record of the result to the database.