Place the required weight on the pan,
Calibration process starts automatically after placing the adequate
weight that is signalled by the following message:
The completion of the calibration procedure is signalled by the
following message:
Take off the weight from the pan, the message <donE> is displayed
for 1s and the program returns to the calibration submenu:
Calibration process can be terminated anytime by pressing
which is signalled by the following message on the display:
Return to weighing with saving changes that have been made.
Caution:
If the calibration process (span adjustment) lasts longer than 15 the
<Err8> message will be displayed and short audible signal will be
emitted. Press
to perform calibration again with more stable
ambient conditions!
- 52 -