20.4.11. Switch off continuous transmission in current unit
Syntax: CU0 CR LF
Possible answers:
CU0_I CR LF
- command comprehended but cannot be executed
CU0_A CR LF
- command comprehended and executed
20.4.12. Send all implemented commands
Syntax: PC CR LF
Possible answers:
PC_- >_Z,T, TO,S,SI,SU,SUI,C1,C0,CU1,CU0,PC – command executed,
the indicator have sent all the implemented commands.
20.5. Manual printouts / automatic printouts
Users can general manual or automatic printouts from the scale.
Manual printouts can be performed after loading the pan
and stabilizing indication by pressing
.
Automatic printouts can be performed only after loading the pan
and stabilizing indication.
Notice:
If a scale is verified printouts of immediate values are blocked.
Format frame:
1 2 3 4 -12 13 14 15 16 17 18
stability space sign mass space unit CR LF