Troubleshooting
R&S
®
RT-ZC31
37User Manual 1801.4949.02 ─ 01
A current exceeding 0.5 A rms has been detected.
If you observe this, implement the remedy described in Table 4-1. Then repeat
the demagnetization and automatic zero-adjustment, see Chapter 3.2, "Demag-
netizing and Zero Adjustment", on page 21.
If demagnetization or automatic zero-adjustment is not completed normally, even
with the sensor not clamped around any conductor and the upper jaw locked in
place, the device can be damaged. Send the device for repair.
The waveform outputted during the demagnetization is asymmetric along
the horizontal axis
This does not represent a device malfunction. After demagnetization and auto-
matic zero-adjustment are completed, make sure that the zero position on the
waveform measuring instrument is appropriate.