Command reference
R&S
®
ZNA
1485User Manual 1178.6462.02 ─ 20
Example:
SYST:COMM:AKAL:CONN THR, 1, 2
Connect a through standard between ports 1 and 2 of the cal
unit.
Usage: Setting only
SYSTem:COMMunicate:AKAL:MMEMory[:STATe] <Boolean>
Shows or hides the memory of the active calibration unit (see SYSTem:
COMMunicate:RDEVice:AKAL:ADDRess).
Setting parameters:
<Boolean> ON - memory is shown in a separate drive.
OFF - memory is not shown.
*RST: OFF
Example:
SYST:COMM:AKAL:MMEM ON
Show the memory of the active calibration unit.
Usage: Setting only
SYSTem:COMMunicate:CODec <Codec>
Selects the character encoding used at the remote interface. The selected encoding
applies to directory and file names, CalKit names, CalUnit characterizations and dis-
play titles.
Parameters:
<Codec> ASCii | UTF8 | SJIS
ASCii: 8-bit ANSI (default)
UTF8: UTF-8
SJIS: Shift JIS
Manual operation: See "Remote Encoding" on page 786
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <GBIBAddress>
Sets the GPIB address of the analyzer.
Parameters:
<GBIBAddress> GPIB address (integer number)
Range: 0 to 30
Example:
SYST:COMM:GPIB:ADDR 10
Set the GPIB address to 10.
*RST; :SYST:COMM:GPIB:ADDR?
After a reset, the address is maintained (the response is 10).
Manual operation: See "GPIB Address" on page 814
SCPI command reference