Command reference
R&S
®
ZNA
1486User Manual 1178.6462.02 ─ 20
SYSTem:COMMunicate:GPIB[:SELF]:DCLear:SUPPress <Boolean>
Suppresses/unsuppresses of Device Clear GPIB interface messages (DCL, SDC).
Parameters:
<Boolean>
Manual operation:  See "GPIB Address" on page 814
SYSTem:COMMunicate:GPIB[:SELF]:INIT:WAIT <Boolean>
Determines/queries the execution behavior of INITiate[:IMMediate] commands
(see Chapter 7.3.8, "INITiate commands", on page 1170).
If set to ON, an automatic *WAI is added (see Chapter 7.2, "Common commands",
on page 882), i.e. the commands execute synchronously.
Parameters:
<Boolean> *RST:  OFF
Manual operation:  See "Advanced ..." on page 815
SYSTem:COMMunicate:GPIB[:SELF]:LPORt:ALIGn <Boolean>
Configures/queries the logical port creation.
If set to ON, logical ports are aligned and must be set from low to high port, which was
the only possibility prior to firmware V1.90. If set to OFF, new ports can be created
freely, like in manual operation. See the example below.
Parameters:
<Boolean> *RST:  ON
Example:  Starting with a 4-port analyzer's default logical port assignment
L1<-->P1, L2<-->P2, L3<-->P3, L4<-->P4,
the remote command
SOUR:LPOR1 1,2
by default (ALIGn=ON) generates aligned logical ports
L1<-->P1&P2, L2<-->P3, L3<-->P4.
With ALIGn=OFF, the result is
L1<-->P1&P2, L3<-->P3, L4<-->P4
and L2 does not exist.
Manual operation:  See "Advanced ..." on page 815
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator <Terminator>
Sets the receive terminator of the analyzer. The receive terminator indicates the end of
a command or a data block.
SCPI command reference