Command reference
R&S
®
ZNA
1512User Manual 1178.6462.02 ─ 20
Manual operation: See "Password" on page 811
SYSTem:PRESet[:DUMMy]
Performs a preset, whose scope is defined using the SYSTem:PRESet:SCOPe com-
mand. The SYSTem:PRESet:USER[:STATe] command determines whether a factory
preset or a user-defined preset is performed.
Example:
SYST:PRES:SCOP SING
Define the scope of a preset: the active recall set is reset only.
SYST:PRES
Reset the parameters of the current recall set to their factory
defaults.
Manual operation: See "Preset Scope" on page 787
SYSTem:PRESet:MODE <PresetMode>
Allows you to select between different preset modes.
GLOBal and RFOFf presets can only be performed from the analyzer GUI. FILE pre-
sets can also be performed via remote commands, if SYSTem:PRESet:REMote[:
STATe] is set to ON.
Parameters:
<PresetMode> NORMal | FILE | GLOBal | RFOFf
NORMal
Factory preset
FILE
User preset file, see SYSTem:PRESet:USER:NAME
on page 1515
GLOBal
Factory preset + GUI reset + reset RF connection configuration
between VNA and external devices
RFOFf
Factory preset + OUTPut<Ch>[:STATe] OFF
Manual operation:
See "Normal / RF Off / Normal, GUI, Ext Setup / <File Name>"
on page 843
SYSTem:PRESet:REMote[:STATe] <Boolean>
Defines the behavior of the preset commands *RST and SYSTem:PRESet[:DUMMy]
Parameters:
<Boolean> OFF (0)
*RST and SYSTem:PRESet perform a factory preset.
SCPI command reference