GUI reference
R&S
®
ZNA
335User Manual 1178.6462.02 ─ 20
5.2.6.1 Controls on the Intermodulation tab
After a [Preset], most of the controls on the "Intermodulation" tab are disabled. They
are enabled after the current channel has been set up for an intermodulation measure-
ment (see "Intermod" on page 335).
Intermod
The "Intermod" button opens the intermodulation setup dialog that allows you to set
up the intermodulation measurement:
●
Non-frequency converting
●
Frequency converting
Intermod Product
After the intermodulation measurement has been set up (see "Intermod" on page 335)
this button opens the intermodulation measurements dialog and activates the Intermod
Product tab.
Intercept Point
After the intermodulation measurement has been set up (see "Intermod" on page 335)
this button opens the intermodulation measurements dialog and activates the Intercept
Point tab.
Main Tone
After the intermodulation measurement has been set up (see "Intermod" on page 335)
this button opens the intermodulation measurements dialog and activates the Main
Tone tab.
Noise
"Noise" (at DUT output) is a measure for the noise level and therefore the dynamic
range of the intermodulation measurement. The noise level is measured at the fre-
quency of the lower tone minus half the tone distance.
Meas softtool