Renishaw LP2 - System Delays and Calibration; System Delays; Calibrating a System

Renishaw LP2
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8
LP2 probe moves
System delays
System delays are repeatable to less than 2µs,
and are constant in each direction in which
measurement is taken.
Delays are automatically compensated for,
provided a calibration move is made in the
same direction and at the same velocity as each
measurement move.
Calibrating a system
Calibrate the probe system at a constant
measurement speed in the measurement
direction, to automatically compensate for errors,
in the following circumstances:
1. Before the system is used.
2. When a new stylus is used.
3. If the stylus is bent.
4. To allow for machine thermal growth.
5. Where there is poor shank relocation
repeatability in the machine spindle.