R 17A User’s Manual – Page 65
2.2 Warranty and maintenance
RESONANCE warrants that:
The R17A is free from defects in material and workmanship under normal use and service for a
period of 36 months from the date of device’s installation by authorized personnel to the first
purchaser.
Accessories are free from defects in material and workmanship under normal use and service
for a period of 12 months from the date of device’s installation by authorized personnel to the
first purchaser.
Furthermore, this warranty shall not apply to, and M.R.S. shall not be responsible for, any loss arising in
connection with the purchase or use of any RESONANCE device that has been:
repaired by anyone other than an authorized RESONANCE service representative;
altered in any way so as, in RESONANCE judgment, to affect its stability or reliability;
subject to misuse or negligence or accident, or which has had the serial number altered,
effaced or removed; or improperly maintained; or used in any manner other than in accordance
with the instructions furnished by RESONANCE.
This warranty is in lieu of all other warranties, express or implied, and of all other obligations or
liabilities of M.R.S., and M.R.S. does not give or grant, directly or indirectly, the authority to any
representative or other person to assume on behalf of M.R.S. any other liability in connection
with the sale of RESONANCE products.
The R17A may be repaired only by your dealer or by a service company authorized by
RESONANCE, recommended by your dealer.
Only original RESONANCE spare or consumable parts must be used for service or repair the
R17A.
In order to ensure that your device works correctly, the R17A Audiometer should be checked and
calibrated at least once a year. We recommend to send the R17A to your RESO-NANCE dealer for this
service, including the probe and all of the accessories.
The verification of instrument calibration must be conducted by qualified service personnel authorized
by the manufacturer via appropriate instrumentation. This must provide the level control and the
frequency of all signals emitted by the transducers supplied with the instrument.
Should transducers require replacement, use only parts provided by the manufacturer. Replacement of
transducers will require re-calibration of the instrument.