BiliTx Professional Manual
18
Maintenance Schedule
Follow the schedule and instructions below to keep the BiliTx system
properly maintained. The light has a minimum lifetime of 30,000
hours. The number of hours your device has been used can be found
in the window on the bottom of the device.
Before Each Patient Use As Needed
Check light intensity. The Joey Dosimeter should
be used to measure the light intensity. If the
intensity falls below the minimum standard,
contact Respironics Customer Service.
Clean the Illuminator device and
ber optic panel.
Replace the disposable light panel cover or vest.
Inspect the power cord and plug for damage or
fraying.
Verify light does not illuminate when panel is not
inserted or connected to the IV pole.