Chapter 6 Specications
25
Standards Compliance
This device is designed to conform to the following standards:
IEC 60601-1 General Requirements for Safety of Medical Electrical Equipment •
IEC 60601-2-50 Requirements for the Safety of Infant Phototherapy Equipment •
Electromagnetic Compatibility: EN 60601-1-2, 2nd edition. •
Electrical Requirements
AC Power 100-240 VAC, 50/60 Hz, 1.0 A
Type of Protection Against Electrical Shock Class I Equipment
Degree of Protection Against Electrical
Shock
Type BF Applied Part
Degree of Protection Against Ingress of
Water
Ordinary Protection, IPX0
Mode of Operation Continuous
Disposal
Dispose of this device in accordance with local regulations.