Chapter 2 Warnings, Cautions, and Symbols
7
During phototherapy, the the baby’s water balance may •
become disturbed. Before and during phototherapy,
make sure the baby is properly hydrated and that his or
her body temperature is maintained.
After treatment has begun, the baby’s bilirubin level •
should be measured to make sure therapy is eective.
Adjust therapy location and time as necessary to avoid •
degradation of photosensitive drugs and infusion liquids.
Do not store any drugs in the illuminated area.
The ber optic panel must not be covered by anything •
except with the cover provided. Any other type of cover
will cause a reduction in light intensity. The setup
instructions must be followed exactly.
Varying ambient conditions, such as the ambient •
temperature and/or dierent radiation sources, may
adversely aect the patient. Please refer to your
hospital phototherapy policy and procedure regarding
appropriate ambient conditions.
Cautions
Keep the unit away from any heated surface. •
Do not scratch, damage, or soil the ferrule end of the •
panel. Also, do not place sharp or heavy items on the
panel, this can damage the panel and aect its light
output.
Do not dry the ber optic panel with articial heat. •
If the device is dropped, contact your an authorized •
service representative to inspect for proper operation.
CAUTION
A caution indicates the
possibility of damage to the
device.