BiliTx Professional Manual
22
Light Source
Minimum Irradiance Level
Standard Panel-Light: 30µW/cm
2
/nm
Neonatal Panel-Light: 55µW/cm
2
/nm
Overhead Therapy: When the illuminator is positioned 30 cm (12 inches)
above the neonate: 30 µW/cm
2
/nm
Wavelength
Blue LEDs: Peak between 460 nm and 480 nm
The graph below shows the total spectral irrandiance of the Illuminator device averaged over a
wavelength range. Also included is a calibration curve of the Joey Dosimeter, which was used to
take the measurements of the light.