EasyManua.ls Logo

Respironics V60 - Page 78

Respironics V60
206 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-10
Operation
2. Adjust Max E:
a. Evaluate the patient. Check whether any of these conditions is
true:
The patient says they are getting too much air, pressure, or
volume
The patient is using accessory muscles to actively stop
inspiration
The Max V or Max P limit is reached
The mask leak has suddenly increased
b. If none is true, increase Max E in increments of 2 cmH
2
O/L while
continuing to evaluate the patient’s response.
c. If any is true, decrease Max E by 2 cmH
2
O/L, and re-evaluate.
Repeat to optimize patient comfort.
3. Repeat the process above adjusting Max R, increasing and decreasing
in increments of 1 cmH
2
O/L/s to optimize patient comfort.
4. Repeat adjustment for Max E as needed.
5. Adjust PPV % downward as tolerated.

Table of Contents

Related product manuals