CAUTION The weight of the SK 300 without collecting receptacle and base frame amounts
approx. 50 kg. The device may only be lifted or transported by at least two people.
The SK 300 has a transportation lug (A) at which the device can be lifted and transported by
means of hoists.
Attach the hoist only to the supplied transportation lug (A).
Please insert a piece of cloth, polystyrene or cardboard between the rear edge of the feed
hopper and the sling or hoisting chain, in order to avoid scratches on the feed hopper.
Lift the device slowly and stabilise it to prevent tipping backwards.
CAUTION Only use suitable hoist that is designed for the weight of the device.
If desired, the transportation lug can be removed after installation and the threaded hole can be
sealed with the supplied plastic screw (PS).
4.3 Temperature Fluctuations and Condensation
Damaged components due to condensation
Temperature fluctuations
− The device may be exposed to substantial fluctuations in temperature
during transport. The ensuing condensation can damage electronic
components.
• Wait until the device has acclimatised before putting it into service.
Temporary storage:
Also in case of an interim storage the device must be stored dry and within the specified
ambient temperature range.
4.4 Conditions for the Installation Site
− Installation height: max. 2 000 m above sea level
− Ambient temperature: 5 °C – 40 °C