RIGOL  Chapter 2 Front Panel Operations 
2-80     DG800 User's Guide 
is set to "Ram Data"), and its available choices include "Head Pos", "Trig Pos", 
and "Tail Pos". 
Tap or use the knob to select the channel to be read, and then set the 
parameters of the corresponding channel. Tap Read. At this time, DS1 will enter 
the "Stop" status automatically, and DG800 will read the arbitrary waveform 
data (i.g. the waveform data collected by DS1) automatically. After reading the 
data, the instrument saves them to the internal RAM of the current channel and 
switches to Sequence mode (if the current mode is not Sequence mode) 
automatically. 
 
4.  Use the BNC cable to connect the output of the current channel of DG800 to the 
input of DS2. Turn on the channel output and DG800 will output the arbitrary 
waveform collected by DS1. You can observe the waveform collected by DS1 
from DS2 by setting DS2.