2. Configure the oscilloscope:
a. To enable the channel, perform any of the following operations:
- Click or tap the channel status label at the bottom of the screen to enable
CH1.
- Press the front-panel key to enable the channel.
- In the Vertical menu, select the CH1 tab. Select ON for the Display item to
turn CH1 on.
b. Set the vertical scale of CH1 to 50 mV/div.
3. Turn on Fluke 9500B; set its impedance to 1 MΩ and select the resistance
measurement function. Read and record the resistance measured.
4. Adjust the vertical scale of CH1 to 500 mV/div; read and record the resistance
measured.
5. Turn off CH1. Measure the resistances of CH2, CH3, and CH4 respectively using the
method above and record the measurement results.
5.2.4 Test Record From
Channel Vertical Scale Test Result Limit Pass/Fail
CH1
50 mV/div 0.99 MΩ to 1.01 MΩ
500 mV/div 0.99 MΩ to 1.01 MΩ
CH2
50 mV/div 0.99 MΩ to 1.01 MΩ
500 mV/div 0.99 MΩ to 1.01 MΩ
CH3
50 mV/div 0.99 MΩ to 1.01 MΩ
500 mV/div 0.99 MΩ to 1.01 MΩ
CH4
50 mV/div 0.99 MΩ to 1.01 MΩ
500 mV/div 0.99 MΩ to 1.01 MΩ
5.3 Input Capacitance Test
5.3.1 Specification
Input Capacitance
Specification 19 pF ± 3 pF
Performance Verification Test