Loading...
RIGOL
Programming Guide for DM3058
5-7
7. CALCulate:NULL:OFFSet
Command Format:
CALCulate:NULL:OFFSet{<value> | MINimum | MAXimum}
CALCulate:NULL:OFFSet?{MINimum | MAXimum}
Function:
The command sets or queries the Null offset.
Parameters:
Name
Type
Range of Values
<value>
Consecutive Integer
± 120% max range
Explanations:
Firstly, do enable calculation function (CALCulate:STATeON) and select limit
function (CALCulate:FUNCtionNULL) to run the command.
CALCulate:NULL:OFFSet{<value> | MINimum | MAXimum}: Set the Null offset.
CALCulate:NULL:OFFSet?{MINimum | MAXimum}: Query the Null offset.
MIN is -120% max range of current function, MAX is 120% max range of current
function.
Return Format:
The query returns the NULL offset.
8. CALCulate:AVERage:AVERage?
Command Format:
CALCulate:AVERage:AVERage?
Function:
The command queries the average of all statistical data.
Explanations:
Firstly, do enable calculation function (CALCulate:STATeON) and select limit
function (CALCulate:FUNCtionAVERage) to run the command.
The average can be queried at any time.
Return Format: