RIGOL
© 2008 RIGOL Technologies, Inc.
Programming Guide for DS1000/DS1000CA Series
2-26
Explanation:
This command is to measure the delay between channel 1, channel 2 at the
negative edge.
Returned Format:
Query returns <-1.00e-04, unit is s.
22. :MEASure:TOTal
Command Format:
:MEASure:TOTal{ON|OFF}
Explanation:
This command is to measure all the parameter value.
Query format:
:MEASure:TOTal?{ON|OFF}
Returned Format:
Query returns ON or OFF. The double quotes are not returned.
Example:
:MEASure:TOTalON Set the total measurement function on
:MEASure:TOTal? Query returns ON
23. :MEASure:SOURce
Command Format:
:MEASure:SOURce<source>
Explanation:
This command is to select the channel.
Query format:
:MEASure:SOURce?
Returned Format: