Query Format:
:TRIGger<mod>:SOURce?
Returned Format:
Query returnsCH1, CH2, EXT, EXT5, AC or DIGITAL. The double quotes
are not returned.
Example:
:TRIGGER:EDGE:SOURCECHANnel1 Set Channel 1 as the Edge trigger source
:TRIGGER:EDGE:SOURCE? Query returns CH1
3. :TRIGger<mode>:LEVel
Command Format:
:TRIGger<mode>:LEVel<level>
Explanation:
The command is to set the voltage level of Edge, Pulse Video trigger.
<mode> can be “:EDGE, “:PULSe” or “:VIDEO”;
<level> is -6*Scale ~ +6*Scale;
Scale indicates the current vertical scale. The default unit of level is V/div.
Query Format:
:TRIGger<mode>:LEVel?
Returned Format:
Query returns the set value of <level> in scientific numeric notation.
Example:
:TRIGger:EDGE:LEVel2 Set the Edge trigger level to 2 V
:TRIGger:EDGE:LEVel Query returns 2.00e+00
4. :TRIGger<mode>:SWEep
Command Format:
:TRIGger<mode>:SWEep{AUTO|NORMal|SINGle}

Other manuals for Rigol DS1000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1000 Series and is the answer not in the manual?

Rigol DS1000 Series Specifications

General IconGeneral
BrandRigol
ModelDS1000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals