Explanation:
The command is to set the trigger type. The options of <mode> are :EDGE,
:PULSe, :SLOPe, :PATTern or :DURation.
AUTO: Acquire waveform even no trigger occurred;
NORMal: Acquire waveform when trigger occurred;
SINGle: When trigger occurs, acquire one waveform then stop.
Query Format:
:TRIGger<mode>:SWEep?
Returned Format:
Query returns AUTO”, NORMAL or SINGLE. The double quotes are not
returned.
Example:
:TRIGger:EDGE:SWEepAUTO Set the AUTO” Edge trigger mode
:TRIGger:EDGE:SWEep Query returns AUTO
5. :TRIGger<mode>:COUPling
Command Format:
:TRIGger<mode>:COUPling{DC|AC|HF|LF}
Explanation:
The command is to set the coupling mode to be DC, AC, HF or LF. The
options of <mode> are :EDGE, :PULSe or :SLOPe.
DC: Allow all signals pass
AC: Block DC signals
HF: Reject high frequency signals (Higher than 150 KHz)
LF: Reject DC and low frequency signals (Lower than 8KHz)
Query Format:
:TRIGger<mode>:COUPling?
Returned Format:
Query returnsDC, AC, HF or LF. The double quotes are not returned.
Example:
:TRIGger:EDGE:COUPlingDC Set the coupling mode to be DC

Other manuals for Rigol DS1000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1000 Series and is the answer not in the manual?

Rigol DS1000 Series Specifications

General IconGeneral
BrandRigol
ModelDS1000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals