Programming Guide for DS1000/DS1000CA Series
:TRIGger:EDGE:COUPling? Query returns “DC”
6. :TRIGger:HOLDoff
Command Format:
:TRIGger:HOLDoff□<count>
Explanation:
The command is used to trigger and stabilize complicated waveform, such as pulse
burst. Holdoff time is the oscilloscope‟s waiting period before starting a new trigger.
During Holdoff, oscilloscope will not trigger until Holdoff ends.
The range of <Count> is 100ns~1.5s.
Query Format:
:TRIGger:HOLDoff?
Returned Format:
Query returns the set value of <Count> in scientific numeric notation.
Example:
:TRIGger:HOLDoff□0.0001 Set the time of holdoff to 100 us
:TRIGger:HOLDoff? Query returns 1.000e-04
7. :TRIGger:STATus?
Query Format:
:TRIGger:STATus?
Explanation:
The command is used to query the current status of the oscilloscope. The status
may be RUN, STOP, T‟D, WAIT, SCAN or AUTO.
Returned Format:
Query returns “RUN”, “STOP”, “T‟D”, “WAIT”, “SCAN” or “AUTO”. The double quotes
are not returned.