Programming Guide for DS1000/DS1000CA Series
2-60
The range of <value> is -6*Scale ~ +6*Scale. Scale indicates the current
vertical scale. The default unit of level is V/div.
Query Format:
:TRIGger:SLOPe:LEVelA?
Returned Format:
Query returns the set value of <value> in scientific numeric notation.
Example:
:TRIGger:SLOPe:LEVelA2 Set the upper trigger level as 2
:TRIGger:SLOPe:LEVelA? Query returns 2.000e+00
4.5 :TRIGger:SLOPe:LEVelB
Command Format:
:TRIGger:SLOPe:LEVelB<value>
Explanation:
The command is to set the lower trigger Level B.
The range of <value> is -6*Scale ~ +6*Scale. “Scale” indicates the current
vertical scale. The default unit of level is V/div.
Query Format:
:TRIGger:SLOPe:LEVelB?
Returned Format:
Query returns the set value of <value> in scientific numeric notation.
Example:
:TRIGger:SLOPe:LEVelB-1.5 Set the lower level of the trigger level as -1.5
:TRIGger:SLOPe:LEVelB? Query returns -1.500e+00
Note: Minimum of Level A couldn’t be less than maximum of Level B.