Programming Introduction RIGOL
Programming Guide for DS1000CA Series
2-31
3. :TRIGger:SLOPe:MODE
Command Format:
:TRIGger:SLOPe:MODE <mode>
:TRIGger: SLOPe: MODE?
Function Explanation:
The commands are used to set and query the slope condition. The options of
<mode> are “+GREaterthan”, “+LESSthan”, “+EQUal”, -GREaterthan”,
-LESSthan” or -EQUal”.
Returned Format:
Query returns “+GREATER THAN”, “+LESS THAN”, “+EQUAL, -GREATER
THAN”, -LESS THAN” or -EQUAL.
Example:
:TRIG:SLOP:MODE +GRE Set the slope condition as “+Greaterthan”
:TRIG:SLOP:MODE? Query Returns “+GREATER THAN”
4. :TRIGger:SLOPe:WINDow
Command Format:
:TRIGger:SLOPe:WINDow <count>
:TRIGger:SLOPe:WINDow?
Function Explanation:
The commands are used to set and query the trigger level of the vertical
window.
The options of <count> are “PA, “PB” or “PAB” when slope condition is
“+GREaterthan”, “+LESSthan” or “+EQUal”;
The options of <count> are NA”, NB” or NAB” when slope condition is
-GREaterthan”, -LESSthan” or -EQUal”;
Returned Format:
Query returns “P_WIN_A, “P_WIN_B”, P_WIN_AB”, “N_WIN_A, “N_WIN_B
or “N_WIN_AB”.

Other manuals for Rigol DS1000CA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1000CA Series and is the answer not in the manual?

Rigol DS1000CA Series Specifications

General IconGeneral
BrandRigol
ModelDS1000CA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals