Chapter 1 Disassemble and Assemble RIGOL
MSO1000Z/DS1000Z Service Guide 1-9
To Disassemble the Keyboards, LCD and Digital
Channel Input Port
Figure 1-6 To Disassemble the Keyboards, LCD and Digital Channel Input Port
Part Explanations:
2 screws (torx countersunk head screw M3*6) fixing the left keyboard.
5 screws (torx countersunk head screw M3*6) fixing the main keyboard.
4 screws (torx countersunk head screw M3*6) fixing the LCD.
2 screws (torx countersunk head screw #FM4#-40*5/16) fixing the digital
channel input port (below the LCD).
Disassemble Steps:
1. Remove the 2 screws () fixing the left keyboard using the screwdriver (T10).
2. Remove the 5 screws () fixing the main keyboard using the screwdriver (T10).
3. Remove the left keyboard cable and main keyboard cable connected to the main
board, and take off the left keyboard and main keyboard gently.
4. Remove the 4 screws () fixing the LCD using the screwdriver (T10) and
remove the screen line connected to the main board. Take off the LCD gently.
5. Remove the 2 screws () fixing the digital channel input port using the
screwdriver (T10), and take off the digital channel input port gently. Note that
only MSO1000Z/MSO1000Z-S includes a digital channel input port.
Left
Keyboard
Main Keyboard
LCD
BNC Washer
BNC Nut