Page #24 background imageLoading...
Page #24 background image
RIGOL Chapter 1 Disassemble and Assemble
1-12 MSO1000Z/DS1000Z Service Guide
and take off the shield cover gently.
(4) Remove the screw () fixing the signal source board using the 5mm
hexagon socket, and take off the signal source board gently.
(5) Remove the screw () fixing the main board and front metal plate using
the 5mm hexagon socket.
(6) Remove the 9 screws () fixing the main board and front metal plate using
the screwdriver (T10).
(7) Take off the main board gently.
2. To Disassemble the Main Board of MSO1000Z/DS1000Z
Part Explanation:
10 screws (torx pan head screw assembly M3*6) fixing the main board and
front metal plate (please remove them according to their actual positions).
Disassemble Steps:
(1) Remove the 4 nuts and washers at the BNC ports on the front metal plate
using the BNC socket.
(2) Remove the 10 screws () fixing the main board and front metal plate
using the screwdriver (T10).
(3) Take off the main board gently.

Other manuals for Rigol DS1104Z

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z and is the answer not in the manual?

Rigol DS1104Z Specifications

General IconGeneral
Bandwidth100 MHz
Channels4
Sample Rate1 GSa/s
Waveform Capture Rate30, 000 wfms/s
Vertical Sensitivity1 mV/div to 10 V/div
Time Base Range5 ns/div to 50 s/div
Vertical Resolution8 bits
InterfaceUSB Host, USB Device, LAN, AUX
Display7 inch TFT LCD (800 x 480)
Trigger TypesEdge, Pulse, Video, Slope
Power Supply100-240 V, 45-440 Hz
Memory Depth24 Mpts (Single Channel), 12 Mpts (Dual Channel)