6.3. Diagnostic Errors 
• Check: Service personnel can check this item on site. 
• Return for Service: The instrument must be returned to the 
manufacturer for factory service. 
Error  Description  Resolution 
(E0001)   The power supply voltage is too 
low.  
Check supply 
(E0002)   The power supply voltage is too 
high. 
Check scale / 
cables 
(E0010)   The temperature is outside of 
allowable limits. 
Check location 
(E0020)  Scale build is incorrect.  The 
number of graduations has been 
set too low or too high. 
Fix up scale build 
(E0100)   The digital setup information has 
been lost.  
Re-enter setup 
(E0200)   The calibration information has 
been lost.  
Re-calibrate 
(E0300)   All setup information has been 
lost  
Enter setup and 
calibrate 
(E0400)   The factory information has been 
lost. 
Return for Service 
(E0800)   The EEPROM memory storage 
chip has failed 
Return for Service 
(E2000)   ADC Out of Range Error.  This 
may be caused from a broken 
load cell cable. 
Check 
BUILD:CABLE 
setting.  Check load 
cell cable, wiring, 
etc. 
(E4000)  The battery backed RAM data 
has lost data.  
Re-enter setup 
(E8000)   The FLASH program memory is 
incorrect  
Return for Service 
The E type error messages are additive.  For example if instrument is 
running off batteries and the temperature drops, the battery voltage may 
be too low.  The resulting error messages will be E0011 (0001 + 0010).  
The numbers add in hexadecimal as follows: 
1 - 2 - 3 - 4 - 5 - 6 - 7 - 8 - 9 - A - B - C - D - E - F 
(For example, 2 + 4 = 6, or 4 + 8 = C