35
new LEDs module (initial intensity=100%).
The LEDs Measure procedure should be run after every change of the LED light source.
Note: The xture has to be connected to mains in order to run LEDs Measure procedure.
Pressure Test -
The menu item runs a procedure which checks a leak tightness of the xture housin
g. The
xture has to be connected to mains and the head temperature cannot be higher than 55°C. The pressure test
lasts about 5 minutes.
Example of screens of the pressure test:
First screen
Test OK
Test failed
Errors
The message "Valve Seal Error" means that valve or coil is defective.
Legend
04.02 ......Time (minutes)
47............Head Temperature (°C)
986.66.....Pressure (hPa)
5.13-------Pressure increment