92
Now check that the displayed value is within the permitted range.
Examine the test strip container and locate the “Accu-Chek Active
Control” table.
The table has two rows listed as “1” and “2”, as well as two
columns giving ranges in mmol/L and mg/dL.
A
CCU
-C
HEK
®
Active Control
mg/dL mmol/L
1 36 – 66 2,0 – 3,7
2 131 – 177 7 ,3 – 9,8
6.3 Quality control test procedure
Values are for illustration only.
man_04493168001_02_EnUK.indd 92 1/12/2006, 11:29:14 ðì