48
Meter and Lancing Device Cleaning and Disinfecting
8
What to Clean and Disinfect
The following parts of the meter should be cleaned and disinfected:
The area around slots and openings (do not get any moisture in slots or openings)
The meter display
The entire meter surface
How to Clean and Disinfect the Meter
WARNING
Failure to follow these instructions will damage the meter and
stop it from working properly.
DO NOT clean or disinfect the meter while performing a blood
glucose or control test.
DO NOT get any moisture in slots or openings.
DO NOT spray anything onto the meter.
DO NOT immerse the meter in liquid.
Always use the same product for both cleaning and
disinfecting.
1
Wash hands with soap
and water and dry
thoroughly.
2
Turn the meter o and
wipe the entire meter
surface with a Super
Sani-Cloth. Carefully
wipe around the test
strip slot and other openings.
Make sure that no liquid enters any slot or
opening.
3
A separate Super Sani-Cloth should be
used for cleaning and disinfection. For
disinfecting the meter, get a new cloth
and repeat step 2, making sure the
surface stays wet for 2 minutes.
Make sure that no solution is seen in any
slot or opening.
4
Wash hands with soap
and water and dry
thoroughly.
82052_08100918004.indd 48 6/9/2020 12:08:31 PM