16
Performa Strip
(Contra-Indications)
User Guide
Contraindication to use for Blood Glucose Meters
Dehydration
Neonatal considerations
Shock
Peripheral occlusive disease
Diabetic Keto-Acidosis (DKA)
Hyperosmolar non-ketotic coma (HONK)
Hypotension
Decompensated heart failure NYHA Class IV
*Refer to test strip package insert for further information.
DOH (HN (Hazard (87) 13)
The system is approved for use with Neonatal blood. As a matter
of good clinical practice, caution is advised in the interpretation
of neonatal glucose values below 2.8 mmol/L.
Please follow recommendations for follow up care that have been
set by your institution for critical glucose values in neonates.
Results for neonates exhibiting signs of Galactosemia should be
confirmed by laboratory results.