111
NOTICE
Damage to the instrument
Ensure that no liquid enters the meter
when disinfecting the test strip guide. If
moisture enters the meter, it may cause
malfunction of the instrument.
Do not insert any objects in the test strip
guide. Doing so may damage the electrical
contacts behind the test strip guide.
Remember to keep the meter on a level
surface prior to disinfecting and powered
off. Squeeze off excess disinfecting solution
or blot on a dry paper towel to remove
any excess disinfecting solution before
disinfecting the surface of the meter.
Cleaning and Disinfecting the Meter and the Lancing Device
CoaguChek Vantus Users Manual, version 1.0