Software and Firmware Update
R&S
®
EB510
54Getting Started 4091.7338.02 ─ 01
If the R&S EB510 cannot reboot
Because the Update32 Tool performs its firmware updates over the LAN,
there could be a conflict if another user is running the tool in the same network.
The R&S EB510 might "hang" after a reboot because the bootloader gets
"trapped" by this Update32 on the network. In such cases the best advise is:
unplug the network cable during boot-up and find out who else is running
Update32 on the corporate network.
4.4 GUI Installation for Remote Access
This section explains how to install the Remote GUI for use on a notebook or desk-
top PC.
Internal installation
If you want to install the GUI on the R&S EB510, follow the procedure descri-
bed in chapter 4.2, "Receiver Firmware and GUI Update for EB510 with front
control panel", on page 40.
4.4.1 Preparations
Required files and accessories
The Remote Access GUI can be retrieved from either one of the locations mentioned
at "CD-ROM and Downloads" on page 40.
Retrieving the IP address of a R&S EB510
It is more convenient to use DHCP with the serial number of the R&S EB510.
For example, if the serial number is 100.001-002, the DHCP name will be "rs-
eb510-100001-002". If DHCP is unavailable, it is advisable to run the
"Update32" tool as explained in: chapter 4.5, "Changing the IP
Address", on page 59 and assign a static IP address.
GUI Installation for Remote Access