19Getting Started 1330.8073.02 ─ 03
3. Select the "Start Self-Alignment" button in the "Alignment" dialog box.
Once the system correction values have been calculated successfully, a mes-
sage is displayed.
To display the alignment results again later
Press the [SETUP] key.
Press the "Alignment" softkey.
Performing a selftest
You do not have to repeat the selftest every time the instrument is switched on. It
is only necessary when instrument malfunction is suspected.
1. Press the [SETUP] key.
2. Select the "Service and Support" softkey.
3. Switch to the "Selftest" tab in the "Service" dialog box.
4. Select the "Start Selftest" button.
Once the instrument modules have been checked successfully, a message is
displayed.
4.1.7 Checking the Supplied Options
The instrument can be equipped with both hardware and firmware options. To
check whether the installed options correspond to the options indicated on the
delivery note, proceed as follows.
1. Press the [SETUP] key.
2. Press the "System Config" softkey.
3. Switch to the "Versions + Options" tab in the "System Configuration" dialog
box.
A list with hardware and firmware information is displayed.
4. Check the availability of the hardware options as indicated in the delivery
note.